escription
1.0 purpose:
1.1 to lay down the procedure for calibration of hplc.
2.0 scope.
2.1 this sop is applicable for the calibration of hplc’s in qc / ipqc laboratories.
3.0 responsibility
3.1 executive-qc
4.0 accountability
4.1 asst. manager-qc
5.0 procedure
5.1 follow the respective sop for operation and carry out the following tests for the calibration of a hplc system as applicable.
5.1.1 flow rate accuracy
5.1.2 gradient composition acuracy
5.1.3 injection volume precision
5.1.4 oven temperature/sample chiller temperature
accuracy
5.1.5 wavelength accuracy
5.1.6 precision /carry over
5.1.7 response linearity
5.2 flow rate accuracy:
5.2.1 flow rate accuracy shall be checked for 0.5 ml, 1.0ml and 2.0 ml/minute with
the following chromatographic conditions.
column: restriction capillary
mobile phase: hplc grade water
5.2.1.1 program the flow rate to 0.5 ml/minute and allow to stabilize.
5.2.1.2 take a clean and dried measuring cylinder of 10ml capacity
and weigh accurately(w1).
5.2.1.3 take a calibrated stop watch.
5.2.1.4 introduce the capillary outlet into the measuring cylinder and
start the stop-watch simultaneously.
5.2.1.5 stop the watch after collection of about 10 ml water in the
cylinder and withdraw the capillary outlet simultaneously.
note the time elapsed as ‘t’.
5.2.1.6 weigh the measuring cylinder containing water accurately
(w2). calculate the weight of water collected (w).
5.2.1.7 calculate the volume (v) of water using the specific gravity of
water at specified temperature as given below. calculate the
flow rate as follows;
flow rate (ml/minute)= v/t
5.2.1.8 repeat steps 5.2.1.1 to 5.2.1.7 and determine the flow rate accuracy for 1.0 ml/min and 2.0 ml/min.
5.2.1.9 record the details as per format no. xxxxx
table 1
volume of 1 g of water at various temperatures
s.no |
temperature( 0c) |
volume (ml) |
1 |
20 |
1.0027 |
2 |
22 |
1.0033 |
3 |
24 |
1.0037 |
4 |
26 |
1.0044 |
5 |
28 |
1.0047 |
6 |
30 |
1.0053 |
5.3 gradient composition accuracy
5.3.1 determine the gradient composition accuracy using the following chromatographic conditions.
5.3.1.1 for low pressure gradient system
pre-purge
wavelength: 265 nm
flow: 2.0 ml/min
stop-time: 10.00
solvent a: hplc grade water.
solvent b:0. 5% acetone in hplc grade water.
timetable:
time solvent b
0.01 100.0
8.0 100.0
8.01 0.0
gradient run:
wavelength: 265 nm
solvent a: hplc grade water.
solvent b:0. 5% acetone in hplc grade water.
flow: 2.0 ml/min
stop-time: 25 + 10 minute’s pre-purge.
table2
time table
time |
solvent b (%) |
1.00 |
0.0 |
1.01 |
100.0 |
5.00 |
100.0 |
5.01 |
90.0 |
8.00 |
90.0 |
8.01 |
50.0 |
11.00 |
50.0 |
11.01 |
6.0 |
14.00 |
6.0 |
14.01 |
5.0 |
17.00 |
5.0 |
17.01 |
50.0 |
20.00 |
50.0 |
20.01 |
0.0 |
25.00 |
0.0 |
5.3.1.1 prior to test gradient composition perform a pre-purge.
5.3.1.2 perform the test in triplicate and measure the average height of every step
5.3.1.3 re-scale the height and noise from absorbance to composition (%b) values.
5.3.1.4 calculate the composition accuracy for every step as the absolute
difference between the rescaled measured step height and the corresponding
gradient composition values.
acceptance criteria : ± 1.500%
5.3.1.2 for high pressure gradient system
pre-purge
wavelength: 254 nm
flow: 1.0 ml/min
stop-time: 10.00
solvent a: hplc grade water.
solvent b:0. 3% acetone in hplc grade water.
timetable:
time solvent b
0.01 10.0
10.0 10.0
gradient run:
wavelength: 254 nm
solvent a: hplc grade water.
solvent b:0. 3% acetone in hplc grade water.
flow: 1.0 ml/min
stop-time: 50 min.
table3
time table
time |
solvent b (%) |
0.01 |
10.0 |
10.00 |
10.0 |
10.01 |
50.0 |
20.00 |
50.0 |
20.01 |
90.0 |
30.00 |
90.0 |
30.01 |
100.0 |
40.00 |
100.0 |
40.01 |
0.0 |
50.00 |
0.0 |
5.3.1.1 prior to test gradient composition perform a pre-purge.
5.3.1.2 perform the test in triplicate and measure the average height of every step
5.3.1.3 re-scale the height and noise from absorbance to composition (%b) values.
5.3.1.4 calculate the composition accuracy for every step as the absolute
difference between the rescaled measured step height and the corresponding
gradient composition values.
acceptance criteria : ± 1.0%
5.4 injection volume accuracy
5.4.1 injection volume precision shall be determined for 10?l, 20 ?l, 50 ?l and 100?l injections.
chromatographic conditions:
mobile phase: hplc grade water.
column: restriction capillary.
run time: 1 min
flow: 1 ml/min
5.4.1.1 fill four vials with hplc grade water and cap them with septum.
5.4.1.2 identify the vials and accurately weigh each of the vials separately.
5.4.1.3 place the vials in the sample tray of the hplc system and program an
injection sequence as follows.
table3
vial injection volume repetitions
vial1 10?l 5
vial2 20?l 5
vial3 50?l 5
vial4 100?l 5
5.4.1.4 after completion of the injection cycle weigh the vials accurately and
determine the average injection volume for each nominal volume using the
volume of water per gram at specified temperature as included in table 1.
record the details as per format no. xxxxx
acceptance criterion: + 10.0% of programmed value.
note : injection volume accuracy is applicable only to hplc system
having autosampler / autoinjector.
5.5 oven temperature/sample chiller temperature accuracy
5.5.1 oven temperature accuracy shall be checked at 300c, 400c and 500c.
5.5.2 sample chiller accuracy shall be checked at the temperatures 100c, 150c and 200c.
5.5.3 set the temperature of the oven /sample chiller to one of the above
mentioned values, insert the temperature probe and allow to stabilize.
after stabilization monitor the temperature at intervals of 5 minutes and
note down three readings of the observed value. repeat the procedure for the other nominal values and record the details as per format no. xxxxx
acceptance criterion: + 20c of programmed value.
5.6 wavelength accuracy
5.6.1 wavelength accuracy of uv detector
chromatographic conditions:
sample: 20 ?g/ml caffeine in water.
run time: 7 min.
wavelength program:

5.6.1.1 fill the flow cell with caffeine standard solution by injecting
directly into the detector inlet and start the program.
5.6.1.2 record the peak response at various wavelengths as per
format noxxxxx
acceptance criterion: wavelength maximum at 273+ 1nm.
note : if the system has no wavelength programming option, than perform the test by selecting wavelength manually for each injection.
5.6.1 wavelength accuracy of pda detector
5.6.1.1 fill the flow cell with caffeine standard solution by injecting directly into the detector inlet and start the wavelength scan from 190 to 400nm.record the details as per format no. xxxxx
acceptance criteria: maxima at 205+ 1 nm and 273 + 1 nm.
5.7 injection precision / carry over
5.7.1 perform injection precision and carry over as follows.
chromatographic conditions;
column: restriction capillary
flow: 1.00 ml/min
run time: 1.00 min
mobile phase: hplc grade water.
injection volume: 20.0?l
wavelength:273 nm
5.7.1.1inject 20.0?l of mobile phase three times followed by six 20?linj. of
caffeine (20?g/ml) in water and again inject one blank (mobile phase).
5.7.1.2 calculate the injection precision for area as the % rsd of caffeine
standard peak responses.
5.7.1.3 calculate the injection carry over for area as peak response in the end blank injection with the peak response from the third blank injection subtracted as % of the response of the sixth caffeine injection. record the calibration details as per format no. xxxxx
acceptance criteria: precision : rsd nmt 2.0%
carry over area: nmt 0.20%.
5.8 response linearity:
5.8.1 determine the detector response linearity by injecting caffeine standard solutions of concentrations 10?g/ml, 20?g/ml, 30?g/ml, 50?g/ml and 100?g/ml in water.
chromatographic conditions: same as 5.7.1
5.8.1.1 inject caffeine standard solutions of the above said concentrations.
5.8.1.2 record the chromatogram and take the peak areas of responses as per
format no. xxxxx
5.8.1.3 plot a linearity graph of concentration vs peak response areas and record
the details as per format no. xxxxx
acceptance criterion: correlation coefficient nlt 0.999
5.9 frequency:
5.9.1 once in six months.
Due to modular nature of the
system, the operation of each
unit is checked properly.
Operation check on overall system
o
Confirm that the system
controller and work station
control each unit during
analysis and that the analysis
results meet the prescribed
criteria.
Software and Firmware check
Here Firmware checking is
conducted based on version display and
Software certificate of Compliance. The
Software and Firmware must be properly
managed and change procedures must be
properly clarified.
Any problems identified in O.Q must be
investigated and appropriate actions must be
taken. All such actions must be documented
and approved by higher authority.
Prior to implementing O.Q, check the
system configuration, determine the items to
be evaluated and record them in O.Q record
and have them approved.
Performance Qualification (PQ):
The objective is to ensure that the
instrument is performing within specified
limits. Hence documented verification that
the equipment and ancillary systems, as
connected together, can perform effectively
and reproducibly based on the approved
process method and specifications.
The PQ represents the final qualification of
equipment or system. This incorporates a
range of testing to simulate your production
process options and provide assurance that
systems and operating documentation are
capable of subsequent process validation
activities. It is used to establish and or
confirm;
1.
Definition of performance criteria and
test procedures.
2.
Selection of critical parameters, with
predefined specifications.
3.
Determination of the test intervals, e.g.,
(a) - Everyday.
(b) - Every time the system is used.
(c) - Before, between and after a series of
runs.
4.
Define corrective actions on what to do if
the system does not meet the established
criteria.
Definition of Calibration: ICH
The demonstration that a particular
instrument or device produces results within
specified limits by comparison with those
produced by a reference or traceable
standard over an appropriate range of
measurements.
Calibration of HPLC:
Various Calibration
parameters are:
Flow rate accuracy
Injector accuracy
System Precision
Wavelength accuracy
Detector linearity
Injector linearity
Gradient Performance Check
Column oven temperature accuracy
Flow Rate Accuracy:
1.
Prime all the solvent lines with Milli Q
water.
2.
Set the flow rate to 0.500 ml/m.
3.
Wait for about 15 m to stabilize the
s
ys
tem and ensure that the pressure is stable.
4.
Insert the outlet tubing into a 10 ml
volumetric flask and start the stop watch
simultaneously.
5.
Stop the stopwatch when the lower
meniscus reaches the 10 ml mark on the
flask.
6.
Record the elapsed time.
7.
Similarly check the flow for 1.0 ml/m
and 2.0 ml/m.
Acceptance criteria:
The time taken to
collect the water should be with in ± 2.0% of
the actual value.
D.
Gowrisankar
et al
/ J Biomed
Sci
and
Res.,
Vol
2 (2),
2010,89-99
90
PQ
Relationships
Table 1:
I.Q And O.Q And P.Q Report
System Name:
System ID No.:
Installation site:
Performer_______________Signature _______________Date:______
Reviewer_______________Signature________________Date:______
Manager_______________Signature________________Date:______
Table 2:
Flow Rate Accuracy
Set Flow
Actual time required to collect
up to the mark in m
Acceptance criteria (in m)
0.5 ml/m
1.0 ml/m
2.0 ml/m
20.0
10.0
5.0
19.6 – 20.4
9.8 – 10.2
4.9 – 5.1
D.
Gowrisankar
et al
/ J Biomed
Sci
and
Res.,
Vol
2 (2),
2010,89-99
91
Table 3:
Chromatographic Conditions for
System Precision
Column
Flow rate
Detection
injection volume
Run time
Column oven temperature
Mobile phase
C
18
,150mm×4.6 mm,5μm
1.0 ml/m
UV at 272 nm
20 μl
15 m
25°C ± 2°C
Water: Methanol (70:30v/v)
Injector Accuracy:
1.
Connect the pump and detector inlet
with union.
2.
Prepare mobile phase consisting of a
mixture of water and Methanol (70:30 v/v)
3.
Set a flow rate of 0.5 ml/m and a run
time of 1 m.
4.
Set the column temperature at 25±
2°C.
5.
Fill a standard HPLC vial to 2/3
rd
with
Milli-Q water. Seal the vial properly with a
cap.
6.
Weigh the vial and record the weight
as W
1
grams.
7.
Place the vial in the chromatographic
system and perform 6 injections of 50μl
volume from this vial.
8.
Weigh the vial again and note the
weigh after the injections as W
2
grams.
Calculate the mean volume injected per
injection as follows:
Mean injected volume (μl) = (W
1
– W
2
)
×100/6
Acceptance criteria:
The mean injected
volume should be 50.0±1.0 μl.
System Precision:
Standard Preparation
: Accurately weigh
and transfer about 60mg of Caffeine into a
100ml volumetric flask. Dissolve and dilute
to the volume with mobile phase. Transfer
10ml of this solution into a 100ml
volumetric flask and dilute to the volume
with mobile phase.
Procedure:
Inject blank, followed by
standard preparation in 6 replicates. Note
down the areas and retention times.
Now calculate the %RSD of retention time
and peak areas for 6 replicates injections.
Acceptance criteria:
The %RSD of
retention time & peak area should be <1.0%.
Wavelength Accuracy:
Procedure:
Create and instrument method
with a wavelength in nm and inject blank,
followed by Standard preparation and note
down the height or absorbance.
Acceptance criteria:
The maximum
absorbance should be ±2nm.
PDA Detector Accuracy:
Select 3D mode and set the wavelength
range as 200-400nm.Inject 20 μl of standard
preparation once into the chromatographic
system. Extract and record the
chromatograms at wavelengths of 202 to
208nm with an interval of 1nm and at 269 to
275 nm with an interval of 1nm.Note down
the height or absorbance.
Acceptance criteria:
The maximum
absorbance should be at 205±2nm and
272±2nm.
D.
Gowrisankar
et al
/ J Biomed
Sci
and
Res.,
Vol
2 (2),
2010,89-99
92
Detector Linearity:
Standard Preparation:
Accurately weigh
and transfer about 60mg of Caffeine into a
100ml volumetric flask. Dissolve and dilute
to the volume with mobile phase.
Detector linearity solution 1(0.06
mg/ml):
Transfer 10ml of Standard
Preparation into a 100ml volumetric flask
and dilute to the volume with mobile phase
Detector linearity solution 2(0.048
mg/ml):
Transfer 8ml of Standard
Preparation into a 100ml volumetric flask
and dilute to the volume with mobile phase.
Detector linearity solution 3(0.03
mg/ml):
Transfer 5ml of Standard
Preparation into a 100ml volumetric flask
and dilute to the volume with mobile phase.
Detector linearity solution 4(0.24
mg/ml):
Transfer 4ml of Standard
Preparation into a 100ml volumetric flask
and dilute to the volume with mobile phase.
Detector linearity solution 5(0.012
mg/ml):
Transfer 2ml of Standard
Preparation into a 100ml volumetric flask
and dilute to the volume with mobile phase.
Procedure:
Inject blank, followed by
Detector linearity solutions and record the
peak responses of Caffeine standard plot
between the concentration Vs the peak
responses.
Acceptance criteria:
The plot should be
linear and regression coefficient (R
2
) should
not be less than 0.99.
Injector Linearity:
Standard Preparation:
Accurately weigh
and transfer about 60mg of Caffeine into a
100ml volumetric flask. Dissolve and dilute
to the volume with mobile phase.
Transfer 10ml of Standard Preparation into a
100ml volumetric flask and dilute to the
volume with mobile phase.
Procedure:
Inject 5 μl of the mobile phase
as blank injection.
Inject 5 μl, 10 μl, 20 μl, 50 μl and 80 μl of
the Standard Preparation and record the peak
areas. Plot a curve for the volume injected
Vs peak area.
Acceptance criteria:
The plot should be
linear and regression coefficient (R
2
) should
not be less than 0.99.
Gradient performance check:
Add 5ml of acetone to 1000ml of methanol
filter and degas. Connect the pump and
detector inlet with union. Set the detector
wave length at 254 nm.
Place Channels A and C in methanol and
channel B and D in 0.5% acetone in
methanol. Set binary gradient with a total
flow rate of 2.0ml/m.Set gradient program
as shown below for channels A, B and C, D
individually.
Table 4:
Gradient Performance Check:
Purge all the channels at a flow rate of
2ml/m for about 5 m. Set the flow rate at
2.0ml/m and wait until the base line is
stable. Set the gradient profile for A and B
Time
(m)
A or C
(Methanol)
B or D (0.5%
Acetone in
Methanol)
Initial
4.00
4.01
8.00
8.01
12.00
12.01
16.00
16.01
20.00
100.0
100.0
90.0
90.0
100.0
100.0
75.0
75.0
100.0
100.0
0.0
0.0
10.0
10.0
0.0
0.0
25.0
25.0
0.0
0.0
D.
Gowrisankar
et al
/ J Biomed
Sci
and
Res.,
Vol
2 (2),
2010,89-99
93
and run the gradient profile by injecting
“0.0”volume of methanol.
Record the height of the peaks. Consider the
height of the peak resulting from B at 100%
concentration as 100 and calculate the
percentage height of other peaks. Perform
the gradient performance check similarly for
channels C and D.
Consider the height of the peak resulting
from D at 100% concentration as 100 and
calculate the percentage height of other
peaks.
Calculations:
Height (%) of B/D =Height of B/D peak
×100/Height of full scale peak
Acceptance criteria:
The calculated
percentage composition (Height (%) should
be with in±1.0% of the set composition.
Column Oven Temperature Accuracy:
It is evaluated with a calibrated digital
thermometer at 30°Cand 60°C .Place the
thermometer probe in the column oven and
set the column oven temperature at
30°C.Wait till the temperature stabilizes.
Record the temperature displayed on the
thermometer. Similarly performs the column
oven temperature accuracy test at 60°C.
Acceptance criteria:
The resulting oven
temperature from the thermometer display
should be within ±2°C of the set
temperature.
NOTE:
a)
For oven Temperature Accuracy,
Chromatographic conditions and mobile
phase refer to system precision test.
Calibration of Gas Chromatogrphy:
Various Calibration parameters are:
Flow rate accuracy
Column oven temperature accuracy
System precision
System precision for head space auto
sampler
Detector linearity
Detector noise and drift test
Flow rate accuracy:
1.
Connect the digital flow meter to the
detector outlet port.
2.
Set the carrier gas (Helium) flow and
wait till it reaches the set flow.
3.
Note the observed flow in replicate.
4.
Repeat the procedure for other carrier
gases such as Hydrogen and Air.
5.
Record the result in GC calibration
protocol.
Acceptance criteria:
The flow rate of
carrier gas should be
±10% of set flow.
Table 5:
Flow Rate Accuracy:
S.No. Carrier
gas
Acceptance
criteria in
ml/m
1. Helium
125
2. Hydrogen
40
3. Air
400
Column Oven Temperature Accuracy:
1.
Connect the column to the detector port.
2. Place the thermometer probe in the
column oven and set the column oven
temperature at 40°C.Wait till the
temperature stabilizes.
3.
Note the observed temperature as read
by the probe in triplicate over a period of 10
m.
4.
Repeat the procedure for 100°C, 150°C
and 190°C.
Acceptance criteria:
The resulting oven
temperature from the thermometer display
should be within ±2°C of the set temperature
System Precision:
Preparation of Standard solution:
Transfer 20 ml of Methanol, Ethanol and
Acetone into 100ml volumetric flask and
make up with Ethyl acetate
Procedure:
Inject blank, followed by
Standard preparation in 6 replicates. Note
down the areas and Retention times.
D.
Gowrisankar
et al
/ J Biomed
Sci
and
Res.,
Vol
2 (2),
2010,89-99
94
Table 6:
Chromatographic Conditions for System Precision
Column 30m
× 0.32mm,1.8μ,DB-624
Detector
Flame ionization detector
Injector temperature
180°C
Detector temperature
250°C
Flow mode
Pressure
Carrier Gas flow rate Helium
25 kpa
Oven program
50°C(hold 5 m) raise to 10°0C
Split ratio
1:10
Injection volume
0.2 μl
Hydrogen flow
40 ml/m
Air flow
400 ml/m
Table 7:
Chromatographic Conditions For Head Space Auto Sampler
Column 30m × 0.32mm,1.8μ,DB-624
Detector Flame
ionization
detector
Injector temperature
220°C
Detector temperature
260°C
Flow mode
Pressure
Carrier Gas flow rate Helium 25 kpa
Oven program
40°C(hold 5 m) raise to 200°C(hold 5 m)
Split ratio
1:10
Injection volume
0.2 μl
Hydrogen flow
40 ml/m
Air flow
400 ml/m
Table 8:
Head Space Conditions
Vial equilibrium
22 m
Vial pressure
0.5 m
Loop fill
0.5 m
Loop equilibrium
0.05 m
Inject 1.00
m
GC cycle time
38 m
Oven temperature
80°C
Loop temperature
100°C
Vial pressure
10.8 psi
Acceptance criteria:
The %RSD of
retention time should be not more than
1.0%& peak area should be not more than
5.0%.
System precision for head space auto
sampler:
Preparation of standard solution:
Prepare
a standard mixture solution by taking
Methylene dichloride (0.6g), Chloroform
D.
Gowrisankar
et al
/ J Biomed
Sci
and
Res.,
Vol
2 (2),
2010,89-99
95
(0.06g),Trichloroethane (0.08g),1,4-Dioxane
(0.38g) in 50ml volumetric flask containing
about 40ml of Dimethyl formamide.Finally
makeup to volume with DMF(Solution-A).
Procedure:
Take 0.5 ml of standard
solution-A in 6 different vials and seal with
septum, then magnetic caps and crimp. Place
these vials on head space sampler; prepare a
blank vial also. Load the vials in head space
sampler tray. Blank vials followed by the
standard vials.
Acceptance criteria:
The %RSD of
retention time should be NMT 1.0%& peak
area should be NMT 15.0%.
Detector linearity:
Preparation of standard solutions:
Detector linearity solution A:
Transfer
10ml of each Methanol, Ethanol and
Acetone into a 100ml volumetric flask and
dilute to the volume with Ethyl acetate.
Detector linearity solution B:
Transfer
15ml of each Methanol, Ethanol and
Acetone into a 100ml volumetric flask and
dilute to the volume with Ethyl acetate.
Detector linearity solution C:
Transfer
20ml of each Methanol, Ethanol and
Acetone into a 100ml volumetric flask and
dilute to the volume with Ethyl acetate.
Detector linearity solution D:
Transfer
25ml of each Methanol, Ethanol and
Acetone into a 100ml volumetric flask and
dilute to the volume with Ethyl acetate.
Detector linearity solution E:
Transfer
30ml of each Methanol, Ethanol and
Acetone into a 100ml volumetric flask and
dilute to the volume with Ethyl acetate.
Procedure: Inject blank, followed by
Detector linearity solutions and record the
peak responses .Draw a standard plot
between the concentrations Vs the peak
responses.
Acceptance criteria:
The plot should be
linear and regression coefficient (R2) should
not be less than 0.99.
Detector Noise and Drift Test:
After GC is ready run the system up to 15 m
through single run. After completion of run
calculate noise and drift through software.
Acceptance criteria:
Noise NMT: 100 μV
Drift NMT: 2500 μV/hr
Calibration of UV-Visible
spectrophotometer:
Spectral calibration:
visible spectral region
Ensure-the socket of the power cord of
the instrument is inserted properly
-cuvettes are clean
Switch ON the instrument. Allow 15m
to warm up.
Keep dummy cuvette in position of
sample holder.
Set the
λ
to 485 nm and press %T
button.
Press 0%T in appropriate direction to
adjust 0.00 reading on read out.
Now remove dummy cuvette from
sample holder. Close the lid.
By adjusting coarse and fine control set
a reading of around 80.0 on read out
No
w set the value of wavelengths in
increm
ents of 0.1 nm up to
λ
of 487 nm and
read the value of %T at each increment of
λ
Draw a curve %T Vs
λ
.
If the peak value of %T is occurring at a
λ
486.1 ± 0.5 nm, the spectral calibration of
the instrument in the visible spectral region
is proper.
This can be confirmed by repeating the
above steps with a maximum value of %T of
around 30.0 on the read out and
λ
setting
from 655 to 657 nm.
If the maximum %T is obtained at a
λ
656.2 ± 0.5 nm, the spectral calibration of
the instrument in the visible spectral region
is confirmed to be proper.
Spectral calibration:
U.Vspectral region:
Keep blank (distilled water) filled
cuvette and sample (benzene vapor) filled
cuvette.
D.
Gowrisankar
et al
/ J Biomed
Sci
and
Res.,
Vol
2 (2),
2010,89-99
96
Table 9:
Chromatographic Conditions For Detector Linearity
Column 30m ×0.32mm,1.8μ,DB-624
Detector
Flame ionization detector
Injector temperature
180°C
Detector temperature
250°C
Flow mode
Pressure
Carrier Gas flow rate Helium
25 kpa
Oven program
50°C(hold 5 m) raise to 100°C
Split ratio
1:10
Injection volume
0.2 μl
Hydrogen flow
40 ml/m
Air flow
400 ml/m
Set the
λ
to 253 nm and press
absorbance button.
Adjust blank to 0.000 on the read out by
using coarse and fine adjustment
Now place the sample into optical path,
value of Absorbance of sample at the
λ
set
appears on the read out.
Again set the values of wavelengths
increments of 0.1nm up to a
λ
of
255nm.Measure the A at each increment. If
maximum A is obtained at
λ
253.9 ± 0.5nm
the “spectral calibration” of the instrument
in U.V region is confirmed to be proper.
Photometric Calibration:
Absorbance:
Visible region
Place dummy cuvette in sample holder
and set %T to “zero”. Now remove dummy
cuvette, by using fine & coarse control set a
reading exactly 40.0 on the read out.
Press Absorbance push button. If the
maximum absorbance obtained at
λ
of
485nm is 0.398 ± 0.002, the photometric
calibration of instrument is confirmed to be
proper.
To confirm, repeat above steps, and set
10.00 on read out
Press Absorbance botton.If the
λ
at 485
nm is 1.000±0.002 then it is confirmed the
photometric performance in the visible
region is proper.
Absorbance
: U.V region
Place blank0.1N H
2
SO
4
cuvette and
60ppm K
2
Cr
2
O
7
as sample
Set
λ
exactly to 257 nm, if the value of
Absorbance of sample at the set
λ
is
0.864±0.005, the instrument is measuring
Absorbance properly.
%Transmittance:
As the value of %T is delivered from
Absorbance itself, if the instrument is
measuring Absorbance properly it is deemed
that it measures %T properly.
Concentration:
Place blank 0.1N H
2
SO
4
cuvette and 60
ppm K
2
Cr
2
O
7
(0.06006g/l of 0.1NH
2
SO
4
) as
standard and 20 ppm (0.02002 g/l of 0.1N
H
2
SO
4
) as sample.
Press “Concentration” push button and
adjust Concentration control to 600 for
standard on read out.
Now place sample holder into optical
path, if the value o f Concentration
appearing on the read out for sample is
200±5, the instrument is measuring
“concentration” properly.
Calibration of Wavelength:
a)
Holmium filter:
For routine calibrations,
holmium filter is satisfactory. Record the
absorption spectrum from 500 to 230 nm
using slowest scan speed and narrowest slit
setting. Identify 3 fused absorption bands
D.
Gowrisankar
et al
/ J Biomed
Sci
and
Res.,
Vol
2 (2),
2010,89-99
97
Table 10:
Chromatographic Conditions
Column 15m ×0.53mm,3.0μ,DB-1
Detector Flame
ionization
detector
Injector temperature
150°C
Detector temperature
200°C
Flow mode
Pressure
Septum purge flow
3 ml/m
Oven program
60°C
Split ratio
1:10
Hydrogen flow
40 ml/m
Air flow
400 ml/m
Run time
15 m
centered on 452.2nm and single band at
around 360.9nm.Instruments with accurately
calibrated
λ
scales will show
λ
max at 453.2,
418.4, 360.9, 287.5, 279.4 and 241.5nm.
b
) Holmium per chlorate solution:
Prepare
a solution of Holmium (III) per chlorate by
dissolving 0.5g of holmium oxide in 2.4 ml
perchloric acid (72% AR grade) by warming
gently and diluting to 10ml with water.
Record the absorption spectrum from 500-
230nm.The wavelengths of principal bands
(Absorbance- 0.4) should be 485.8, 450.8,
416.3, 361.5, 287.1, 278.7, 241.1 nm.
c)
Discharge lamps: A
low pressure
discharge lamp is suitable. Record the
transmission spectrum from 600 to 240 nm
of Mercury lamp place near the entrance to
monochromator, using minimum slit setting
and slowest scan speed. The principal
emission lines of Mercury are at 579.0,
576.9, 546.1, 435.8, 404.5, 364.9 and 253.7
nm.
d) Prepare standard solution by dissolve 100
mg of Potassium dichromate in 0.05N
Potassium hydroxide solution in 100ml
volumetric flask. Make up to volume with
the same. From the standard solution take
20ml and make up to 500ml with 0.05N
Potassium hydroxide solution. Now scan the
wave length from 340 to 400nm using blank
0.05N Potassium hydroxide solution. The
maximum wave length is observed at 370
nm.
Limit of stray light:
Weigh accurately 1.2g of dried Potassium
chloride in 100 ml volumetric flask and
makeup to mark with Double distilled water.
Measure the absorbance at 200 nm.
Acceptance criteria:
Tolerance limit NLT
2.0
Resolution:
Prepare 0.02%v/v solution of Toluene and
make up with Hexane. Scan the wavelength
from 250 to 280nm.Maximum absorbance is
269 nm and Minimum absorbance is 266nm
Acceptance criteria:
Ratio limit NLT 1.5
Photometric linearity:
Weigh accurately 100mg of Potassium
chromate in 100ml volumetric flask and
dissolve in 0.05N Potassium hydroxide
solution. Make up with the same solvent.
From the above solution take 20ml and
make up to 500ml with 0.05N Potassium
hydroxide solution.
Now prepare dilution of 4,8,16,24,32
μg/ml
Measure the absorbance at 370nm using
blank.
Acceptance criteria:
The plot should be
linear and regression coefficient (R2) should
NLT 0.999.
D.
Gowrisankar
et al
/ J Biomed
Sci
and
Res.,
Vol
2 (2),
2010,89-99
98
Table 11:
Schedule for Calibration/Inspection of Some Major Instruments
INSTRUMENT INTERVAL (MONTHS)
HPLC 3±7
days
Gas Chromatography
3±7days
UV-Visible spectrophotometer
Monthly once ±3days
IR spectrophotometer
Monthly once ±3days
NMR spectrophotometer
6 ±15days
Flourimeter 3±7
days
Polarimeter
Monthly once ±3days
PH meter
daily
Analytical balance
daily
Abbreviations:
HPLC : High Performance Liquid
Chromatography
GC : Gas Chromatograph
ICH : International Conference on
Harmonization
C18 : Octadecyl silane column
DMF : Dimethyl formamide
DB-624: 6% Cyano Propyl Phenyl-
94%Dimethyl Polysiloxane
NMR : Nuclear Magnetic Resonance
IR : Infra red
The calibration of the entire analytical
instrument or its components (which ever is
appropriate) should be performed after any
major maintenance.
Conclusion:
During all phases of clinical
development, including the use of small-
scale facilities or laboratories to
manufacture batches of APIs for use in
clinical trials, procedures should be in place
to ensure that equipment is calibrated, clean
and suitable for its intended use.
Procedures for the use of facilities should
ensure that materials are handled in a
manner that minimizes the risk of
contamination and cross-contamination. So
validation and calibration is very important
for analytical instruments.
References:
[1]
Lloyd R.Snyder,Joseph J.Kirkland, Joseph
L.Glanjch.Practical,
HPLC method
development, A Wiley
– interscience
publication,1997.
[2]
USP 30 NF 25 “The official compendia of
standards”, Asian edition, 2007, vol: 1.
[3]
A.H.Beckett, J.B.Stenlake,
Practical
Pharmaceutical chemistry
, CBS Publishers
and distributors, IV-Edition, part two, 2004,
p.no.326.
[4]
Y.Anjaneyulu, R.Marayya,
Quality Assurance
and Quality Management in Pharmaceutical
industry
, Pharmabook Syndicate, 2005.
[5]
ICH Guidelines Q2BValidation of Analytical
Procedures: Methodology (1996
)
No comments:
Post a Comment