Operation And Calibration Of HPLC SYSTEM
1.0. OBJECTIVE:
The objective of this SOP is:
1.1 To ensure that the instrument performs satisfactorily and gives accurate and reproducible data.
2.0. RESPONSIBILITY:
2.1 Officer or above of Quality control -
3.0. ACCOUNTABILITY:
Head - Quality Control
4.0. PROCEDURE:
4.1 PROCEDURE FOR GENERAL CLEANING:
Ø Ensure that the power supply to the instrument is switched off and main cord is removed from supply.
Ø Clean the instrument with a clean dry cloth everyday. A wet cloth dipped in dilute soap solution may be used occasionally.
Ø Precaution must be taken to clean the instrument immediately with dry cloth.
4.2 OPERATING INSTRUCTIONS:
Ø Ensure that the instrument is properly connected to the power supply.
Ø Fix the column and prepare the mobile phase.
4.3 PREPARATION OF MOBILE PHASE
Ø Use HPLC grade solvent / water & AR/ HPLC grade reagent only.
Ø Filter the mobile phase through 0.45 membrane filter after mixing it in required proportion and de-gas on ultrasonic bath for about 5 minutes.
4.4 PROCEDURE TO GET STARTED
Ø Switch on pump, auto injector, detector and then system controller.
Ø Turn the drain valve knob to 180 in anti- clockwise to open the drain valve to run the purge system.
Ø After the purge is over close the drain valve.
Ø Set desired flow rate by pressing function key.
Ø Press pump key. The pump will run and indicator will glow.
Ø Purge the auto injector by pressing purge key on auto sampler display.
Ø Enter the desired wavelength on detector using function key
4.5 PROCEDURE TO OPERATE CLOSE UP
Ø Switch on the computer monitor and printer.
Ø Double click on the main Menu of software which is you use for your hplc system .
Ø It will show Confrigration,Real Time ,Sample Shedule & Post Run Analysis.
Ø Open Real Time.
Ø From the “file menu” create a new method or load the existing method.
Ø Create a new sample shedule and feel sample name ,batch number,,sample volume , method name and file name ,after then save the shedule .
Ø Run Sample Shedule .
Ø For system suitability test, inject six continuous injections of same standard and RSD should not be more than 2.0 %, otherwise it is specified in standard test procedure.
Ø After every ten injection of sample (5 samples in duplicate), the standard solution shall be injected 3 times and the RSD shall be calculated and ensure that it is within the limit.
4.6 SHUT DOWN PROCEDURE
Ø Close Sample Shedule ,Real Time and then CLASS LC 10 to Software.
Ø From Start button select “ Shut down” and click yes.
Ø Switch-off the computer.
Ø Switch-off CBM .
Ø Switch-off Detector then, auto sampler.
Ø Stop pump by pressing, “Pump” key and switch off the pump module.
4.7 CALIBRATION PROCEDURE.
4.7.1 FOR PUMP:
Ø Disconnect the column and connect the inlet and outlet tubing’s with a union.
Ø Prime all the lines at 5 ml/min flow rate with water and ensure that flow line is free from air bubbles.
Ø Set the flow rate at 1ml / min and collect the mobile phase (water) in a dry preweighed beaker and collect the mobile phase for 10 min. Weigh the beaker to get the weight of mobile phase.
Ø Calculate the flow rate by dividing the weight obtained with weight per ml and 10 (run time).
Ø Calculate the corresponding flow rate. Carry out the experiment in duplicate.
Ø Repeat the same procedure for Pump- B
Ø Record the observation in Annexure -1
Ø Acceptance criteria : Flow rate should be in between 0.99 to 1.01 ml / min..
4.7.2 FOR GRADIENT VALVE:
Ø Install union in place of column & flush solvent lines (A&B) at flow rate of 2ml/min with water.
Ø Prepare the mobile phase.
Ø Prepare 0.3% acetone with HPLC grade water.
Ø Fill reservoir A with 100% HPLC grade water & reservoir B with 0.3% acetone in HPLC grade water as mobile phase.
4.7.3 INSTRUMENT SET UP:
Ø Enter the following time program:
TIME
|
FUNCTION
|
VALUE
|
0.01
|
B CONC
|
10
|
10.00
|
B CONC
|
10
|
10.01
|
B CONC
|
50
|
20.00
|
B CONC
|
50
|
20.01
|
B CONC
|
90
|
30.00
|
B CONC
|
90
|
30.01
|
B CONC
|
100
|
40.00
|
B CONC
|
100
|
40.01
|
B CONC
|
0
|
50.00
|
B CONC
|
0
|
Ø Use detector at wavelength of 254 nm.
Ø Record the printout of gradient valve test as per Annexure - 2.
Ø The gradient valve test shall be accepted if actual concentration with ±1% of set concentration.
4.7.4 CALIBRATION OF INJECTOR:
4.7.4.1 CHECK FOR PRECISION:
Ø Purge the injector system with 100% water to ensure the complete washing of injector.
4.7.5 STANDARD PREPARATION:
Ø Transfer about 50 mg of Uracil to a 250ml volumetric flask. Add 100ml of Methanol, sonicate to dissolve and make up the volume with Methanol to obtain a solution containing about 0.02 mg/ml of Uracil.
Ø Filter the solution through 0.45m membrance filter.
Ø Use HPLC grade Methanol as the mobile phase.
Ø Instrumental Set Up
Column : Hypersil ODS or equivalent 150 x 4.6 mm, 5.0m
Flow rate : 1.0 ml/min
Injector Volume : 20m litre
Detector : 254 nm
Ø Inject the standard preparation six times in the system. The peak areas observed shall be consistent.
Ø The relative standard deviation for area counts calculated shall not be more than 1.0%.
Ø Record the observation in the format as mentioned in Annexure – 3.
4.7.6 CHECK FOR LINEARITY:
Ø Inject 10, 20, 30, 40 & 50m litre of standard preparation in duplicate. Calculate the average area counts corresponding to each set of injection.
Ø Tabulate the average area against each injection. Plot a graph for area counts vs mlitre the resulting graph shall be linear.
Ø The correlation co- efficient calculated shall be not less than 0.99.
Ø Record the observation as per the Annexure - 3.
4.7.7 CALIBRATION OF DETECTOR:
Ø Standard preparation. Mobile phase, Instrument set- up same as mentioned in calibration of Injector.
Ø Run the chromatograph at different wavelength (252 nm – 262 nm with 2 nm increment)
Ø The largest peak response shall be at 258 nm ± 2nm.
Ø Record the results in the detection calibration record as per the Annexure - 4.
Ø Affix a calibration status label on the instrument containing “Calibrated On”, “Due On” and “Signature”.
Report to Head – QC, if any discrepancy observed during calibration or operating the instrument and affix ‘Under Maintenance’ label on the instrument
FREQUENCY OF CALIBRATION:
Ø Detector : Once in three months and after each maintenance job.
Ø Pump : Once in three months and after each maintenance job.
Ø Injector : Once in six months and after each maintenance job.
Ø Gradient : Once in six months and after each maintenance job in the pump.
Your article has piqued a lot of positive interest. I can see why since you have done such a good job of making it interesting.
ReplyDeleteautocampionatore spazio di testa